Chapter 6 Reprocessing, Storage, and Disposal
HIGH FLOW INSUFFLATION UNIT UHI-4
Disinfection
Follow the usage (temperature, contact time, use life) and
amount (concentration) provided by surface disinfectant
cleaner manufacturer. Failure to follow manufacturer’s
instructions may result in insufficient cleaning and
disinfection effect.
Wipe off remaining surface disinfectant cleaner solution
according to the instructions of the manufacturer of the
surface disinfectant cleaner. Failure to do so may adversely
affect the human body or this instrument.
After wiping with a moistened lint-free cloth, dry the High
Flow Insufflation Unit and Smoke Evacuation Foot Switch
thoroughly before using it again. If it is used while still wet,
there is the risk of an electric shock.
1. Turn OFF the High Flow Insufflation Unit and disconnect the power cord
from the hospital grade wall mains outlet.
2. Prepare a lint-free cloth moistened with the surface disinfectant cleaner.
Figure 6.3
3. Wipe off all patient debris from the High Flow Insufflation Unit and Smoke
Evacuation Foot Switch using the moistened lint-free cloth.