HIGH FLOW INSUFFLATION UNIT UHI-4
Guidance and manufacturer’s declaration —
Electromagnetic immunity
This model is intended for use by medical personnel in hospital
environments and for use in the electromagnetic environment specified
below. The customer or the user of this model should assure that it is used
in such an environment.
This instrument can be used with the high-frequency electrosurgical
equipment that designated by Olympus.
Immunity test
IEC 60601-1-2 (2014)
test level
IEC 60601-1-2
(2007, 2001)
test level
Compliance
level
IEC 60601-1-2 (2007, 2001)
Electromagnetic environment
— Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact:
8 kV
Air:
2, 4, 8, 15 kV
Contact:
2, 4, 6 kV
Air:
2, 4, 8 kV
Same as left Floors should be made of wood,
concrete, or ceramic tile that
hardly produces static. If floors
are covered with synthetic
material that tends to produce
static, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV
for power supply lines
1 kV
for input/output lines
2 kV
for power supply
lines
1 kV
for input/output
lines
Same as left Mains power quality should be
that of a typical commercial
(original condition feeding the
facilities) or hospital
environment.
Surge
IEC 61000-4-5
Differential mode:
0.5, 1 kV
Common mode:
0.5, 1, 2 kV
for signal input/
output lines: 2 kV
Differential mode:
0.5, 1 kV
Common mode:
0.5, 1, 2 kV
Same as left Mains power quality should be
that of a typical commercial or
hospital environment.