Important Information — Please Read Before Use
HIGH FLOW INSUFFLATION UNIT UHI-4
When using the port device for transanal access in
combination with the high flow insufflation unit, carefully read
and understand the following before using:
Anytime you suspect an abnormality in the high flow
insufflation unit, immediately stop using the high flow
insufflation unit.
Over-insufflating the lumen may cause patient pain, injury,
bleeding, gas embolism, and/or perforation.
Be sure that this instrument is not used adjacent to or
stacked with other equipment (other than the components of
this instrument or system) to avoid electromagnetic
interference.
Electromagnetic interference may occur on this instrument
near equipment marked with the following symbol or other
portable and mobile RF (Radio Frequency) communications
equipment such as cellular phones. If electromagnetic
interference occurs, mitigation measures may be necessary,
such as reorienting or relocating this instrument, or shielding
the location.
Before using a veress needle, trocar, and port device for
transanal access, inspect the items as outlined in Section
4.10, “Inspection of UHI-4 operation when using a veress
needle/trocar/port device for transanal access”.
This instrument requires consideration of electromagnetic
compatibility (EMC), as well as installation and initial setup
procedures in accordance with this manual.
Do not touch the electrical contacts inside the high flow
insufflation unit’s connectors. Otherwise, equipment damage
and/or malfunction can occur.
Do not use a pointed or hard object to press the buttons on
the front panel. This may damage the buttons.
Do not apply excessive force to the connectors. Otherwise, a
failure of an electrical contact may result in a malfunction.
Turn the high flow insufflation unit OFF before connecting or
disconnecting cables. Equipment damage or malfunction can
result.