THB/SO2-8 Operator’s Manual – OPTI CCA-TS2
tHb/SO
2
ANALYTES
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
tHb
Parameter and Specication Value
Number of Samples 45
Range 7.7 to 21.2
Slope 1.01
Offset -0.09
R
2
0.996
Sy|x 0.298
6
8
10
12
14
16
18
20
22
6 8 10 12 14 16 18 20 22
Linear fit