Operator’s Manual – OPTI CCA-TS2 THB/SO2-9
ANALYTES tHb/SO
2
SO
2
Parameter and Specication Value
Number of Samples 63
Range 71.5 to 99.9
Slope 1.03
Offset -3.54
R
2
0.992
Sy|x 0.939
References
1. Tietz, Norbert W., Ed., ClinicalGuidetoLaboratoryTests, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Tietz; Burtis C, et al (Eds.), TextbookofClinicalChemistryandMolecularDiagnostics, 4
th
Ed.,
(Elsevier Saunders, 2006) pps. 2252-2302.
3. Tietz; Burtis C, et al (Eds.), TextbookofClinicalChemistryandMolecularDiagnostics, 4
th
Ed.,
(Elsevier Saunders, 2006) p. 1004.
4. OPTI Medical. Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).
5. NCCLS. ReferenceandSelectedProceduresfortheQuantitativeDeterminationofHemoglobinin
Blood;–ApprovedStandard 3
rd
Edition; NCCLS document H15-A3. NCCLS, Wayne, PA, 2000.
65
70
75
80
85
90
95
100
65 70 75 80 85 90 95 100
SO2 - TS2
SO2 - TS
Scatter Plot with Fit
Identity
Linear fit
(-3.54 + 1.03x)