3-6 Operator’s Manual – OPTI CCA-TS2
3 CUSTOMIZATION
Option 2:
<QC Lockout Enable> (Fig. 3-13).
When this option is enabled, external QC
measurements must be performed at specied
intervals for patient measurements to be allowed.
1. To specify the number of QC measurements to
be performed, make sure <Lockout By Level
Enable> is not selected and select 1, 2 or 3 in
the <By Number> option.
2. To specify the levels to run, select <Lockout
By Level Enable> and the levels to run in the
<By Level> option.
3. Dene the time interval during which QC
measurements must be run. Options are
8, 12, 24 hours,7 days and 1 month.
4. By default, the analyzer is set to run 3 levels of
external controls at 1 month intervals.
5. To disable QC lockouts, deselect <QC Lockout
Enable>.
6. Press to accept the changes.
NOTE: Theselectedtimeintervalstartswiththe
timethisfeatureisactivated.
NOTE: Morethanoneoptioncanbeselected.
Forinstance,laboratoriescanrequirethat
acombinationofSRCsandliquidQCisrun
onadailybasis.Thisshouldbebasedon
hospitalpolicy.
NOTE: Controllockoutsarebasedondatastoredin
theControlsdatabase(seeSection4).
Thisdatabasemayincludedatameasured
withanycassettelotorcassettetype.
Fig. 3-13 QC Lockout