REF : M18-SM-001 Rev 02
5. OUTPUT FORMAT
MYTHIC 18 Copyright© Orphée SA. All Rights Reserved. Page 30/68
5. OUTPUT FORMAT
The choice of the setting (format, speed …) is explained in page N° 24 of the User’s Manual.
5.2 C2 FORMAT
5.2.1 General principle
Mythic begins frame with headline
MYTHIC X ;Y ;Z( ;eventual parameters)[CR]
Where :
X is the number of the Mythic (maximum 2 characters).
Y is the user identification (login) (maximum 10 characters).
Z frame identification.
The decimal separator is the dot (.).
The field separator is « ; ».
The line separator and the indicator of end is CR.
The uncut frame are :
Request of results send with send results.
Send calibration pack with results of calibration.
All the lines of a frame must be sent. ( If no information, the parameter is not fulfilled).
All the identifiant (frame or parameters) are in block letters.
All editable text field (Identification, lot) are coded in UTF8 ( non Latin languages management).
5.2.2 FRAMES
5.2.2.1 REQUEST TO SEND RESULTS
Mythic sends results after acknowledgement by the host.
Frames used in « Hand-Shake » mode.
5.2.2.1.1 Request to send results
MYTHIC X ;Y ;RESULT_READY ;Size[CR]
5.2.2.1.2 A
CKNOWLEDGEMENT REQUEST TO SEND RESULTS
ACK_RESULT_READY[CR]
5.2.2.1.3 A
CKNOWLEDGEMENT RESULTS
ACK_RESULT;A;B[CR]
Where A : OK or Error Code, B : Reserved.

Table of Contents

Other manuals for Orphee MYTHIC 18