9. REPARING
REF : M18-SM-001 Rev 02
Page 56/68 Copyright© Orphée SA. All Rights Reserved. MYTHIC 18
9.3 COUNTING BATHS MODULE
9.3.1 Baths dismantling
Perform baths dismantling as described in the User’s Manual.
9.3.2 Baths o-ring replacement
Perform baths o-ring replacement as described in the User’s Manual.
9.3.3 Aperture block replacement
Perform aperture block replacement as described in the User’s Manual.
9.3.4 Complete baths module replacement
- Before to replace the complete syringe module, drain all the
tubing :
x Remove the diluent, lysis and cleaner straw from their
bottle
x Perform a PRIME ALL (MAIN MENU / REAGENT / PRIME
ALL)
- Switch off the MYTHIC and disconnect the power supply.
- Remove the Lysis and Cleaner bottle from their location
- Then put the MYTHIC on its left side.
Remove the baths connection (HGB, WBC coaxial, RBC coaxial
and thermal captor).
Unscrew (but
WITHOUT dismantle them) the three screws
which maintain the baths module (the third screw is located
behind the over flow tubing).
Cut the fixation ties of the tubing.
Disconnect the tubing from the manifold and the bath.
Slide up and remove the baths module.
NOTA : Take care about the tubing reconnection, if necessary, cut few millimeters (2 to 3mm) of tubing or
exchange them.
REASSEMBLY PROCEDURE :
To reassemble the unit, carry out the various operations in the opposite direction.