REF : M18-SM-001 Rev 02
5. OUTPUT FORMAT
MYTHIC 18 Copyright© Orphée SA. All Rights Reserved. Page 33/68
5.2.2.2.3 RBC pathologic messages
ANE
ERY>
MICRO
MACRO
MICR>
MICR>>
MICR>>>
MACR>
MACR>>
MACR>>>
ANIS>
ANIS>>
ANIS>>>
HYPOCR
COLDAGG
NO_INTERPRETATION
5.2.2.2.4 PLT pathologic messages
THR>
THR<
MACROP
PLTAGGR
MICROC
SCHIZ
CELLD
NO_INTERPRETATION
5.2.2.3 CALIBRATION
MYTHIC X;Y;CALIBRATION;Y ;Calibration date ;calibration time ;LOT ;Expiry date ;Creation date ;Creation
time ; ;X ;A ;B ;C ;D ;F ;Number of results [CR]
WBC; target value ; tolerance [CR]
RBC; target value ; tolerance [CR]
HGB; target value ; tolerance [CR]
HCT; target value ; tolerance [CR]
PLT; target value ; tolerance [CR]
END_CALI ; Checksum [CR]
Where
X : User who creates the lot.
Y : User who performs the calibration.
A, B, C, D and F are the coefficient of the calibration, for respectively, WBC, RBC, HGB, HCT and PLT.
Follow N results which served for calibration.
In case of manual input of the coefficients, information of the lot are empty and the number of results is
null.
Acknowledgement of the calibration is done by the frame
ACK_CALI ;Lot;A [CR]
Where A : OK or Error Code.