8. MAINTENANCE
The quality of the results and the reliability of the MYTHIC 18 are directly linked to the strict respect of
the maintenance hereafter described.
To perform the maintenance and the repair described in this section, it is mandatory to wear
rubber gloves and wash hands with a disinfectant after completion of work.
8.1 MAINTENANCE TABLE
NOTA : This table is valid for an average number of 50 samples per day. For more, please increase
proportionally the number of maintenances.
MAINTENANCE
DAILY WEEKLY MONTHLY HALF A YEAR ANNUALLY
User Tech User Tech User Tech User Tech User Tech
Reagents level X
Start Up X
Automatic cleaning X
Concentrate cleaning X
Shut down X
Cover cleaning X
Piston greasing X X
Needle o-ring replacement X
Syringes o-ring replacement X
Motor screw greasing X
- Reagents level control : see User’s Manual.
-
Start up : see User’s Manual.
-
Automatic Cleaning : see User’s Manual.
-
Concentrated cleaning : see User’s Manual.
-
Shut down : see User’s Manual.
-
Cover cleaning : see User’s Manual.
-
Piston greasing : see User’s Manual.
- Needle o-ring replacement : see section 9.2.1
- Syringes o-ring replacement : see section 9.4.2
- Motor screw greasing : This operation is described during the ORPHEE training.