8. MAINTENANCE
MYTHIC 18 Copyright© Orphée SA. All Rights Reserved. Page 53/68
- Wait for the information window.
- Open the door on the right side (see section
1.1.3).
- Put 2 ml of sodium hypochlorite solution in each
counting chamber.
- Press the button in the window located in the middle
of the screen.
- MYTHIC 18 performs a maintenance cycle of the
apertures followed by a stand by mode during 2 min.
- After 2 min. MYTHIC 18 rinses all of the elements.
Operator can perform an analytic cycle.
-
Wear rubber gloves and wash hands with a
disinfectant after completion of work.
8.3 CLEAN OUT
This procedure has to be used when the MYTHIC 18 must be transported for a long distance by car, train or
plane. This procedure consists of cleaning the MYTHIC 18 with sodium hypochloride, then rinsing it with
distilled water and drying it before the transportation.
The clean out is accessible from MAIN SCREEN then SERVICE then
TECHNICIAN.
- Press