8. MAINTENANCE
REF : M18-SM-001 Rev 02
Page 52/68 Copyright© Orphée SA. All Rights Reserved. MYTHIC 18
8.2 CONCENTRATED CLEANING
Such cleaning is necessary when MYTHIC 18 is in permanent rejection for one measured parameter.
Prepare a Sodium Hypochlorite solution at 12° in chloride (diluted 4 times with distilled water if sodium
hypochlorite is concentrated at 48° in chloride).
- Return to the main menu by pressing
.
- Then press
.
- Press
.
- Press
to start the cycle (drain the counting
chamber).