10. TROUBLESHOOTING
REF : M18-SM-001 Rev 02
Page 68/68 Copyright© Orphée SA. All Rights Reserved. MYTHIC 18
MESSAGE ACTION
MECA : HOME ROCKER NOT FOUND Perform an INIT MOTOR or a CONTROL CYCLE.
MECA : HOME NEEDLE NOT FOUND
Perform an INIT MOTOR or a CONTROL CYCLE.
MECA : HOME SYRINGE NOT FOUND
Perform an INIT MOTOR or a CONTROL CYCLE.
MECA : MOTOR NEEDLE GAP Nothing.
MECA : MOTOR ROCKER GAP Nothing.
MECA : MOTOR SYRINGE GAP Nothing.
MECA : NEEDLE NOT IN TOP POSITION Perform a CONTROL CYCLE.
MECA: MOTOR NEEDLE BUSY Re-start the Mythic.
MECA: MOTOR ROCKER BUSY Re-start the Mythic.
MECA: MOTOR SYRINGE BUSY Re-start the Mythic.
NO PRINTER RESPONSE Switch on the printer or invalidate the printings.
NO PRINTER RESPONSE Switch on the printer or invalidate the printings.
NO PRINTER SELECTED Switch on the printer or invalidate the printings.
OUT OF RANGE Modify the value
PRINTER ERROR Switch on the printer or invalidate the printings.
PRINTER IS BUSY Switch on the printer or invalidate the printings.
PRINTER IS OFF Switch on the printer or invalidate the printings.
PRINTER: NO PAPER Add some paper.
RINSE NOT DONE Perform a RINSE ALL cycle.
RS232: ACK ERROR Re-start the Mythic.
RS232: INTERNAL ERROR Re-start the Mythic.
RS232: SYNCHRO ERROR Re-start the Mythic.
RS232: TIME OUT Re-start the Mythic.
RUNNING AUTO CLEANING , Press OK.
START UP CYCLE FAILED Perform a new start up cycle
START UP CYCLE NOT DONE Perform a start up cycle.
SVM: BAD VERSION. Update the SVM software
SVM: COM. TIME OUT Re-start the SVM.
SVM: ILLEGAL SERIAL NUMBER. This MYTHIC can not be connected to the SVM
SVM: UNMATCH
Re-enter the file or confirm it (manual connection on the
SVM).
SVM: WG Westgard alarm.
SVM: XB XB alarm.
SYSTEM : EEPROM COM ERROR Re-start the Mythic.
SYSTEM : INTERNAL TIME OUT Re-start the Mythic.
SYSTEM: FATAL ERROR Re-start the Mythic.
THE CLEANER USED IS OUT OF DATE Replace the bottle and perform a prime Cleaner
THE DILUENT USED IS OUT OF DATE Replace the container and perform a prime Diluent
THE LYSE USED IS OUT OF DATE Replace the bottle and perform a prime Lysis
WASTE ALMOST FULL Replace the waste container