Otoport OAE+ABR
User Manual for Otoport DP+TE model
125
Conguration
15.2.5 DPOAE
The Otoport DP+TE has three dierent DPOAE test modes, DP High, DP
Low and DP 12 Freq. Choose the mode you wish to edit from the menu.
DP Levels
Conguring Levels sets two DP stimulus pairs (L1 and L2), which are
available for selection prior to test. Stimuli in DP High must be of a higher
intensity than those set in DP Low.
Note:
It is not possible to set the DP High mode stimulus settings
lower than the DP Low mode stimuli and vice versa, but they
may be set to equal levels. Stimulus settings are automatically
corrected if they do not obey this condition.