Otoport OAE+ABR
User Manual for Otoport DP+TE model
211
Technical specications
Guidance and manufacturer’s declaration – electromagnetic emissions
The Otoport OAE+ABR is intended for use in the electromagnetic
environment specied below. The user of the Otoport OAE+ABR should
ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment
- guidance
RF emissions Group 1 The Otoport OAE+ABR uses RF energy
CISPR 11 only for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference in
nearby equipment.
RF emissions Class B The Otoport OAE+ABR is suitable
CISPR 11 for use in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
Harmonic emissions Class A purposes.
IEC 61000-3-2
Voltage uctuations/ Complies
icker emissions
IEC 61000-3-3
27.3 Electromagnetic Compatibility:
Otoport OAE+ABR
The Otoport OAE+ABR should be put into service according to the EMC
(Electromagnetic Compatibility) information provided here.
Portable and mobile RF (Radio Frequency) communications equipment can
aect the operation of the Otoport OAE+ABR. In particular, mobile telephones
(‘cellphones’) should not be operated within 10m of the Otoport OAE+ABR.
The use of probes, electrodes, chargers and connection cables other than those
supplied by Otodynamics Ltd, and specically for use with the Otoport OAE+ABR
may result in increased emissions or decreased immunity of the Otoport OAE+ABR.
The Otoport OAE+ABR should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary the Otoport OAE+ABR
should be observed to verify normal operation in the conguration in which it is used.