EasyManua.ls Logo

Otodynamics Otoport Flex - Page 212

Default Icon
228 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER TWENTY-SEVEN
Technical specications
212
Technical specications
Guidance and manufacturers declaration – electromagnetic immunity
The Otoport OAE+ABR is intended for use in the electromagnetic environment specied
below. The user of the Otoport OAE+ABR should ensure that it is used in such an
environment.
Immunity test IEC 60601 Compliance Electromagnetic
test level level environment - guidance
Electrostatic ± 6 kV contact ± 6kV contact Floors should be wood,
discharge (ESD) concrete or ceramic tile.
± 8 kV air ± 8 kV air If oors are covered with
IEC 61000-4-2 synthetic material, the
relative humidity should be
at least 30%.
Electrical fast ± 2 kV for power ± 2kV for power Mains power quality should
transient/burst supply lines supply lines be that of a typical
commercial or hospital
IEC 61000-4-4 ±1 kV for input/ ±1kV for input/ environment.
output lines output lines
Surge ± 1 kV line(s) to ± 1kV di󰀨erential Mains power quality should
line(s) mode be that of a typical
commercial or hospital
IEC 61000-4-5 ± 2 kV line(s) to ± 2kV common environment.
earth mode
Voltage dips, <5% U
T
<5% U
T
Mains power quality should
short interrup- (>95% dip in U
T
) (>95% dip in U
T
) be that of a typical
tions and voltage for 0.5 cycle for 0.5 cycle commercial or hospital
variations on environment. If the user
power supply 40% U
T
40% U
T
of the Otoport OAE+ABR
input lines (60% dip in U
T
) (60% dip in U
T
) requires continued operation
for 5 cycles for 5 cycles during power mains
IEC 61000-4-11 interruptions, it is recommended
70% U
T
70% U
T
that the Otoport OAE+ABR be
(30% dip in U
T
) (30% dip in U
T
) powered from an uninterruptible
for 25 cycles for 25 cycles power supply or a battery.
<5% U
T
<5% U
T
(>95% dip in U
T
) (>95% dip in U
T
)
for 5 s for 5 s
Power frequency 3 A/m 3 A/m Power frequency magnetic
(50/60 Hz) elds should be at levels
magnetic eld characteristic of a typical
location in a typical
IEC 61000-4-8 commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.

Table of Contents

Related product manuals