58
58
It is critical that the seating system is closely monitored during the automatic calibration.
If any strange noises (actuators binding) are heard during the automatic calibration, TURN OFF the
wheelchair immediately. Also, visually confirm that the actuators are moving to both ends of travel during
the calibration sequence. If an actuator “stops in the middle” during automatic calibration, there may
be a problem with that actuator or sensor (try performing a manual calibration to attempt to resolve the
problem.)
Errors during Automatic Calibration
The Automatic Calibration sequence performs several checks to confirm that each of the five calibration
steps have been successful. If the ICS detects that a step was not completed successfully, it will give
notification of this failure by sounding the R-net joystick’s horn continuously. Note how many LEDs are
illuminated on the switchbox and the position of the seating system before turning off the chair to silence
the notification (and exit the calibration mode.) Try automatic calibration a second time, if the failure
occurs again, attempt a manual calibration (explained on the next page.)
Programming K450 - Specific Attributes