Reviewing Trending Data 15 Vital Signs Trending
159
NOTE
When the time interval in the Vital Signs Trending Report is one minute, data from continuous
measurements represents the average reading for that one-minute period. For all other time intervals,
the measurement shown is the one-minute average from the most recent minute in the time interval.
Vital Signs Trending Report Parameter List Order
Vital Signs Trending Reports list parameters in the following order, beginning at the top of the table:
HR, P1, P2, ABP, Ao, ART, PAP, RAP, CVP, LAP, ICP, CPP, NBP, EtCO
2
, AwRR, SpO
2
, Pulse,
Temp.
If a parameter has not been measured during the display period, it is not listed in the display.
Scrolling in the Vital Signs Trending Report
Use the [<<] or [>>] softkeys to scroll left and right (backward and forward) in the Vital Signs
Trending Report. The softkey will be displayed as inactive if there is no more data to be viewed in a
particular direction.
If there are more vital signs than can be shown on the screen, use the Navigation buttons next to the
Menu Select button to scroll up and down with the vertical scroll bar on the display. Make sure there is
not a menu active at the same time.
Vital Signs Trending Report Intervals
Vital Signs Trending data can be shown at selected intervals for up to 12 hours of monitoring. You can
adjust the display’s time interval for the current incident to 1, 5, 10, 15, 30 or 60 minutes. The default
is 5 minutes.
Adjusting Vital Signs Trending Report Interval
To adjust the intervals in the Vital Signs Trending report, perform the following steps:
1 With Vital Signs Trending active on your display, press the Menu Select button.
2 Using the Navigation buttons, select Tre n d I n t e r v a l and press the Menu Select button.
3 Using the Navigation buttons, select the trend interval you want and press the Menu Select button.