Operational Check 19 Maintenance
229
Operational Check Report
The Operational Check takes only a short time to complete. When it is done, a report is printed, as
shown in Figure 86. The first part of the report lists test results. The second part lists checks to be
performed by the user.
Figure 86 Operational Check Report
Press the [Print] soft key when the Operational Check is complete to print an additional copy of
the report.
Operational Check Report
Model Number: M3535A
Serial Number: USD0123456
Ver.: 7.00.00
Current Operational Check:
DD Mon YYYY HH:MM:SS Pass
Last Operational Check:
DD Mon YYYY HH:MM:SS Pass
Current Test Results:
General System Test: Pass Battery Compartment A Test: Pass
Therapy Knob: Pass Battery Compartment B Test: Pass
Charge Button: Pass
Shock Button: Pass
Audio Test: Pass
Defib Test: Pass/Pads
Pacer: Pass
Compression Sensor: Pass
Leads ECG Test: Pass/ECG Cable
Pads/Paddles ECG Test: Pass/Pads
SpO2 Test: Pass
NBP Test: Pass
CO2 Test: Pass
Invasive Pressure Test: Pass
Temperature Test: Pass
Bluetooth Test: Pass
Printer Test: Pass
Qty/Check List:
___ Defibrillator Inspection ___ SpO2 Sensor
___ Cables/Connectors ___ NBP Cuffs & Tubing
___ Paddles/Pads ___ CO2 FilterLine
___ Compression Sensor Inspection ___ Invasive Pressure Cables
___ Compression Sensor Pads ___ Temperature Probes
___ Monitoring Electrodes
___ Charged Batteries
___ AC/DC Power & Cord
___ Printer Paper
___ Data Card
Comments:
Inspected by:_________________