Automated Tests 19 Maintenance
217
Automated Test Summary
An Automated Test Summary (ATS), showing the results of recent tests, may be printed as evidence
that the HeartStart MRx is tested regularly. To display the ATS and print its results:
1 Turn the Therapy knob to Monitor.
2 Press the Menu Select button.
3 Using the Navigation buttons, select Other and press the Menu Select button.
4 Select Operational Check and press the Menu Select button.
5 Using the Navigation buttons, select Automated Test Summary and press the Menu Select button.
The Automated Test Summary is displayed.
The message, Leaving Normal Operating Mode, appears to let you know that you are exiting from
clinical functionality of the monitor/defibrillator.
6 Press the [Print] soft key to print the report.
The report shows the results of the most recent Hourly Test, the Daily Tests that have run since the
last Weekly Test, and the last 53 Weekly Tests. Test results are reported, as described in Table 55.
Table 55 Automated Test Summary Results
Result RFU Indicator Definition Required Action
Pass Hourglass All tests passed None
Fail/
DX
Solid red X, chirp A problem has been detected
that may prevent the delivery of
a shock, pacing, or ECG
acquisition.
Turn the Therapy Knob to Monitor. An
inop indicating a problem has occurred is
displayed. Refer to Chapter 20,
Troubleshooting, for the action to take.
Fail/BF Blinking X The total battery capacity
(combination of both batteries)
is less than 20%.
Charge the battery as soon as possible and/
or replace the battery with a charged
battery. Charging may be done in the
HeartStart MRx or by connecting to AC/
DC power, or in a Philips-approved battery
support system.
Fail/D Hourglass A problem has been detected
with a component that does not
affect therapy delivery.
Turn the Therapy Knob to Monitor. An
inop indicating the failed component is
displayed. Refer to Chapter 20,
Troubleshooting, for the action to take.