16 Q-CPR and Data Capture CPR Feedback
176
CPR Feedback
Adjusting CPR Feedback Volume
To adjust the volume of CPR feedback voice prompts:
1 Press the Menu Select button.
2 Using the Navigation buttons, select Volume from the menu and press the Menu Select button.
3 Select Voice and press the Menu Select button.
4 Select the desired volume level and press the Menu Select button.
A sample voice prompt is annunciated to confirm your selection.
NOTE In AED Mode, steps 2 and 3 are not applicable unless Q-CPR is active.
If desired, CPR feedback voice prompts may also be muted. When muted, a CPR Voice Muted text
message displays in the middle of the wave sector containing the compression waveform during
Manual Defib Mode or below the text prompt in AED Mode. CPR inactivity time is indicated by an
audible tone.
To mute the CPR feedback voice prompts:
1 Press the Menu Select button.
2 Press the Menu Select button again to select Mute CPR Voice.
To resume voice prompts set at the previously selected volume:
1 Press the Menu Select button.
2 Press the Menu Select button again to select Resume CPR Voice.
NOTE Mute is only available when Q-CPR is active.