17 Configuration Configurable Parameters
184
Configurable Parameters
The following tables list configurable parameters for each of the sub-menus on the Configuration Main
menu. A description of each parameter is provided, along with the possible choices for settings. Default
settings are in bold type.
Table 2 4 General Settings
NOTE The Return to Owner Password should be documented in a secure manner. Should it need to be reset,
all configuration settings must be reset to factory default values, and the device must then be
re-configured.
NOTE Do not set alarm volumes so low that you can’t hear them in loud environments.
Parameter Description Setting Choices
Voice Volume Voice pr ompt volume level. Ver y Sof t, Soft , Medium, Loud, Very Loud
Alarm Volume Audible alarm volume level. Very Soft, Soft, Medium, Loud, Very Loud
Minimum Alarm Volume Minimum audible alarm level available during
use.
Very Soft, Soft, Medium, Loud, Very Loud
QRS Volume Volume of audible beep with each QRS
complex detected.
Off, Very Soft, Soft, Medium, Loud, Very Loud
Time Format Selects time format to display. 12-hour, 24-hour
Pacing on Batteries
Warning
Displays the warning message
Pacing on
Batteries
if pacing using only battery power.
Yes, No
Unit Display Defines if parameter values are displayed with
or without corresponding measurement units.
On, Off
Patient Category Selects the default patient category. Adult, Pedi
Device Owner Identification for Return to Owner feature. 20 characters
Return to Owner
Password
Password required in order to deactivate the
loan period setting.
4 characters
Audio Recording Turns audio recording on or off for Manual
Defib or AED modes.
On, Off