16 Q-CPR and Data Capture Using Q-CPR in Manual Defib Mode
170
Using Q-CPR in Manual Defib Mode
When the Therapy Knob is moved to the 150J Manual Defib setting, the CPR sub view is
automatically displayed only if the Sync function is disabled, patient category is Adult, and patient age
is equal to or greater than 8 years old. The CPR sub view is automatically displayed after the Manual
Therapy Security password is entered, only if Manual Defib Mode is password protected.
When the Therapy Knob is located on any other Manual Defib setting, a compression of the
Compression Sensor displays the CPR sub view. The CPR sub view may also be manually displayed by
pressing the [Start CPR] soft key. Also, the patient category must be Adult and the patient age
must be equal to or greater than 8 years old to display the sub view for either action.
Wave forms for ECG, invasive pressure (ABP, ART Ao or PAP) and CO
2
parameters can be viewed
during the use of Q-CPR. If an ABP, ART Ao or PAP pressure measurement is active, the waveform
always appears in Wave Sector 3 and the pressure value displays in Parameter Block 2. If two pressures
labeled ABP, ART, Ao or PAP are available, the label assigned to Pressure 1 is displayed. If no pressure
is active, then the compression waveform appears in Wave Sector 3. Parameter Block 2 displays
compression and ventilation measurement information. The EtCO
2
waveform will appear in Wave
Sector 4 if present. AwRR numeric values are not displayed.
NOTE If the Pads/CPR cable is not connected when you turn the Therapy Knob to Manual Defib Mode, the
message Connect Pads/CPR Cable displays.
NOTE When using Q-CPR, SpO
2
monitoring functionality is not available.
NOTE Compression and ventilation measurement values are printed in the annotation area of the ECG
printed strip.