Specifications 21 Specifications and Safety
287
Environmental (M3536A)
Temperature: 0
º
C to 45
º
C operating; -20
º
to 70
º
C storage
Charging the battery at temperatures above 45
º
C may degrade battery life
Storing the battery for extended periods at temperatures above 40
º
C will reduce battery capacity and
degrade battery life
Humidity: Up to 95% Relative Humidity
Printer paper may jam if paper is wet
Thermal Printer may be damaged if wet paper is allowed to dry while in contact with printer
elements
Altitude:
Operating: 0 to 15,000 ft (0 to 4,500 m)
Storage: 0 to 15,000 ft (0 to 4,500 m)
Shock:
Operating: Half-sine waveform, duration < 3 ms, acceleration > 145 g, 1 time on all six faces
Non-operating: Trapezoidal waveform, acceleration >
30 g, velocity change=742 cm/s ±10% on all
six faces
Bump: EN60068-2-29 Bump (Half-sine, 40 g peak, 6 msec duration, 1,000 bumps x 3 axes)
Free Fall: IEC 68-2-32 Free Fall. Drops on all faces onto a steel surface (excluding bed rail hook)
- 30 in. (76.2 cm) with carrying case
- 16 in. (40.6 cm) without carrying case
Vibration:
Operating: MIL STD 810E 514.4 Category 6 Helicopter, General Storage, UH60
•Non-Operating:
- IEC 68-2-6 Vibration (sinusoidal) (10-57 Hz+
0.15mm; 58-150 Hz, 2g; 20 sweeps x 3 axes)
- IEC 68-2-64 Vibration, broad-band random (10-20 Hz, 0.05 g
2
/Hz; 20-150 Hz, -3 dB/octave;
150 Hz, 0.0065 g
2
/Hz; 1.5 hours x 3 axes)
Solids/Water Resistance: IP24. Water testing performed with cables connected to the device.
EMC: Complies with the requirements of standard EN 60601-1-2:2001.
Safety: Complies with the requirements of applicable safety standards.
Other Considerations:
The HeartStart MRx is not suitable for use in the presence of concentrated oxygen or a flammable
anesthetic mixture with air, oxygen, or nitrous oxide.
Hazards arising from software errors were minimized by the product’s compliance with the software
requirements contained in EN 60601-1-4:1996.