21 Specifications and Safety Specifications
286
Water/Solids Resistance:
Water Resistance: Meets EN60601-2-4. Water testing performed with cables connected to the device.
IPX4 - when used with water resistant paddles (M3543A) the M3535A is protected against splashing
water.
Solids Resistance: IP2X.
EMC: Complies with the requirements of standard EN 60601-1-2:2001.
Safety: Complies with the requirements of applicable safety standards.
Other Considerations:
The HeartStart MRx is not suitable for use in the presence of concentrated oxygen or a flammable
anesthetic mixture with air, oxygen, or nitrous oxide.
Hazards arising from software errors were minimized by the product’s compliance with the software
requirements contained in EN 60601-1-4:1996.
Mode of Operation: Continuous
AC Line Powered: 100 - 240 VAC, 50 - 60 Hz, 1 - 0.46 A (Class 1)
Battery Powered: 14.8 V Rechargeable, Lithium Ion
DC Powered:
Input: 11-32 VDC, 11 A
Output: 18 V, 5 A, 90 W