Specifications 21 Specifications and Safety
281
Pulse Rate Resolution 1bpm
Pulse Rate Accuracy: 2% or 1 bpm (whichever is greater) over the measurement range
Pulse Alarm Range:
Low Limit: 30 to 195 (Adult); 30 to 235 (Pediatric)
High Limit: 35 to 200 (Adult); 35-240 (Pediatric)
Pulse Response Time (90 to 120 bpm): average 18.0 seconds; standard deviation 0.86 seconds
NBP
Pressure Range:
• Systolic: 40-260 mmHg
• Diastolic: 20-200 mmHg
Pulse Rate Range: 30 to 220 bpm
Initial Pressure: 160 mmHg Adult; 120 mmHg Pediatric
Maximum Pressure: 280 mmHg
Overpressure Safety Limits: Maximum of 300 mmHg
Cuff Inflation Time: 75 second maximum (pediatric or adult)
Pressure Transducer Accuracy: ±3mmHg
Alarm Range:
• Systolic high limit: 35-270 (Adult), 35-180 (Pediatric)
• Systolic low limit: 30-265 (Adult), 30-175 (Pediatric)
• Diastolic high limit: 15-245 (Adult), 15-150 (Pediatric)
• Diastolic low limit: 10-240 (Adult), 10-145 (Pediatric)
• Mean high limit: 25-255 (Adult), 25-160 (Pediatric)
• Mean low limit: 20-250 (Adult), 20-155 (Pediatric)
Rated Life: 50,000 measurement cycles (36/day for 2.3 years)
Auto Mode Repetition Time: 1, 2.5, 5, 10, 15, 30, 60, or 120 minutes
Maximum Measurement Time: 120 sec.
Interconnect Tube Length:
• M1598B Connect tubing 1.5 m
• M1599B Connect tubing 3.0 m
Recommended frequency of pressure transducer calibration: yearly