Configurable Parameters 17 Configuration
185
Table 25 Heart Rate and ECG Settings
Parameter Description Setting Choices
Auto Gain Determines whether ECG size is automatically
adjusted to the standard maximum wave size
without clipping the wave sector. If auto-gain is
off, the gain is set to 1 (10mm/mV).
On, Off
AC Line Filter Selects the setting used to filter out AC line
noise. Adjust this setting to the power
frequency of your country.
60 Hz, 50 Hz
ECG Bandwidth for
Display
Selects the display filter frequency for the
attached therapy cable or 3-/5-lead ECG cable.
(This setting does not affect 12-Lead View.)
.15-40 Hz Monitor, 1-30 Hz EMS
ECG Bandwidth for
Printer
Selects the printer filter frequency for the
attached therapy cable or 3-/5-lead ECG cable.
(This setting does not affect 12-Lead Reports.)
If Diagnostic is selected, pads ECG is set to
Monitor.
0.05-150 Hz Diagnostic, .15-40 Hz Monitor,
1-30 Hz EMS
ECG Electrode Labels Selects electrode label format. AAMI: RA, LA,
LL, RL, V.
IEC: R, L, F, N, C.
AAMI, IEC
HR/Pulse High Limit Selects the default High Alarm Limit for the
HR derived from the ECG and the pulse
derived from SpO
2
or Invasive Pressures.
Adult: 35-200, 120, (bpm)
Pediatric: 35-240, 160
adjusted in increments of 5
HR/Pulse Low Limit Selects default Low Alarm Limit for the HR
derived from the ECG and the pulse derived
from SpO
2
or Invasive Pressures.
Adult: 30-High Alarm Limit, 50 (bpm)
Pediatric: 30- High Alarm Limit, 80
adjusted in increments of 5
VTACH Detection of ventricular tachycardia based on
heart rate limit and PVC run limit.
Heart Rate Limit:
Adult: 95-150, 100 (bpm)
Pediatric: 95-150, 120
adjusted in increments of 5
PVC run limit:
Adult: 3-20, 5
Pediatric: 3-20, 5
adjusted in increments of 1