Symptoms 20 Troubleshooting
251
Event Storage Full message. Event Summary data for the
current patient case exceeds the
storage capacity of internal memory
or you’ve reached the 12-hour limit.
• Copy or print Event Summary to retain data and
then begin a new patient case.
• Print the Event Summary Report. Begin a new
patient case.
Power Supply Failure
INOP. There is a problem with the
internal power supply.
Remove the device from use and call for service.
Replace Clock Battery
message.
The battery on the internal clock
needs to be replaced.
Call for service.
Shock Equip Malfunction
INOP.
A shock can not be delivered due to
a hardware failure.
Remove the device from use and call for service.
Shutting Down in 1 Minute
message.
Very low battery and the device is
not connected to AC/DC power.
Insert a charged battery and/or connect to AC/DC
power.
Shutting Down Now message. Battery charge is depleted and the
device is not connected to AC/DC
power.
Insert a charged battery and/or connect to AC/DC
power. Settings for the current patient may need to be
reset.
Table 62 General Problems (Continued)
Symptom Possible Cause Possible Solution