Physio Control LIFEPAK CR2 - No Voice Prompt

Physio Control LIFEPAK CR2
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Title Service Manual, LPCR2 Page
90 of
104
Doc
Type
Service Product Drawing
Doc
ID
3323299
Doc
Rev
B
Doc
State
Released State Date 11/14/2017
7000545 Revision A
Start
CR
2
Readiness light
is not flashing
Open the Lid
Can you hear
voice prompt?
Press Two buttons
Language + Child
Yes
Replace Battery
No
Open the Lid
Can you hear
voice prompt?
Close the Lid
Yes
Replace the Unit
End
No
Follow the voice
prompt direction
Is Readiness light
flashing
Yes
No
2. If flashing, but email notification indicates that CR2 needs attention, use
Using Voice Prompts for
Troubleshooting for troubleshooting.
No Voice Prompt
The possible causes:
Depleted batteryreplace the battery
Speaker system failure or device failureafter replacing the battery, still has no voice prompt, replace the unit