6.3-8
Section 6
ApplicAtions
Pinnacle Operators Manual
Pickering Laboratories Inc.
METHOD 3: 1846150 C
18
COLUMN (4.6X150MM) WITH METHANOLIC SAMPLES
HPLC FLOW RATE: 1.0ML/MIN COLUMN TEMPERATURE: 42˚ C REAGENT FLOW RATE: 0.3 ML/MIN
CONDITIONS FOR METHANOLIC SAMPLES
Time (Min) Water % MeOH %
0 85 15
0.5 85 15
28.5 30 70
28.6 0 100
33.5 0 100
33.6 85 15
41 85 15
Carbamate Test Mix, 2.5ppm, 10µl injection, 15 cm, Carbamate C
18
column(1846150)
1
2
3
4
5
10
10
11
20 30
6
7
8
9
min
0
N