6.3-11
Section 6
ApplicAtions
Pinnacle Operators Manual
Pickering Laboratories Inc.
METHOD 6: 0846250 C
8
COLUMN (4.6X250MM) WITH AQUEOUS SAMPLES
HPLC FLOW RATE: 1.0ML/MIN COLUMN TEMPERATURE: 42˚ C REAGENT FLOW RATE: 0.3 ML/MIN
CONDITIONS FOR AQUEOUS SAMPLES
Time (Min) Water % MeOH %
0 100 0
1 100 0
1.1 82 18
36 30 70
39 30 70
39.1 0 100
41 0 100
41.1 100 0
55 100 0