Chapter A - GENERAL & TECHNICAL DATA
ProMax X-ray unit with Dimax3 A-13
EMC INFORMATION
Technical Manual
7 EMC INFORMATION
WARNING
Use of any accessories and cables other than those specified in Planmeca Pro-
Max X-ray unit’s documentation, with exception of cables sold by Planmeca as
replacement parts for internal components, may result in increased emission
or decreased immunity of the X-ray unit.
WARNING
Planmeca ProMax X-ray unit should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the Planmeca ProMax
X-ray unit should be observed to verify normal operation in configuration
which it will be used.
Guidance and manufacturer’s declaration - electromagnetic emissions
Planmeca ProMax X-ray unit is intended for use in the electromagnetic environment specified
below. The customer or the user of the Planmeca ProMax X-ray unit should assure that it is used
in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
Planmeca ProMax X-ray unit uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
Planmeca ProMax X-ray unit is suitable for use
in all establishments, including domestic estab-
lishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies