Chapter C - ERROR MESSAGES
ProMax X-ray unit with Dimax3 C-13
HELP AND ERROR MESSAGE SHORT-FORM TABLES
Technical Manual
1.8 Calibration errors (7xx)
Error
code
Explanation Comments
E701 Missing calibration Tube head filament definition not done Perform the preheat calibration
again.
E702 Tube head kV-calibration not done Perform the preheat calibration
again.
E703 Panoramic AEC not calibrated Calibrate the AEC-PAN.
E704 Cephalometric AEC not calibrated Calibrate the AEC-CEPH.
E705 Problem during the preheat calibration. The current in the tube head is
not high enough. Perform the
preheat calibration again.
E706 Problem with the preheat calibration. The current in the tube head is
not raising on the voltage area.
Perform the preheat calibration
again. Check the functionality of
the tube head.