Chapter H - CEPHALOSTAT
ProMax X-ray unit with Dimax3 H-29
ADJUSTMENTS AND CALIBRATIONS
Technical manual
1.13 Adjusting the second primary collimator position
Second primary collimator angle adjustment
The second primary slot must be perpendicular to the head support in both directions.
sideways and in depth. Check the collimator position in depth by using the spirit level as
shown on the Fig. 50 below.
Figure 50
Loosen the three attachment screws of the collimator plate with 3mm Allen key, if needed.
Figure 51
PXR_Adj_digiceph30.eps
90°
PXR_Adj_digiceph11.eps
Collimator plate attachment screws