3 EMC INFORMATION
Technical manual Planmeca ProX 15
3 EMC INFORMATION
WARNING
Use of any accessories and cables other than those specified in Planmeca ProX
intraoral X-ray unit’s documentation, with exception of cables sold by Planmeca as
replacement parts for internal components, may result in increased emission or
decreased immunity of the X-ray unit.
WARNING
Planmeca ProX intraoral X-ray unit should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the Planmeca ProX
intraoral X-ray unit should be observed to verify normal operation in configuration
which it will be used.
Guidance and manufacturer’s declaration - electromagnetic emissions
Planmeca ProX intraoral X-ray unit is intended for use in the electromagnetic environment specified
below. The customer or the user of the Planmeca ProX intraoral X-ray unit should assure that it is used
in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
Planmeca ProX intraoral X-ray unit uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Planmeca ProX intraoral X-ray unit is suitable for
use in all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies