4 SERVICE MODE
24 Planmeca ProX Technical manual
4.10 Ready-state setting
The Planmeca ProX intraoral X-ray unit can be set so that the Ready indicator light will
only come on when the Planmeca Romexis program is ready for the exposure, i.e. the
Waiting for exposure
message is on the computer screen.
Enter the service mode according to the instruction given in section 4.2 “How to enter/exit
service mode” on page 19.
Press the parameter adjustment up key until the parameter number 21 appears on the kV
display.
The number indicating the ready-state setting (0 or 1) is shown on the time display.
0: Normal operation (factory setting). The Planmeca ProX intraoral X-ray unit will go into
Ready-state regardless of PC operation.
1: The Ready indicator light will only come on when the Planmeca Romexis program is
ready for the exposure
Press and hold down the Select key until the number starts to blink, and the number can
now be changed with the parameter adjustment keys.
Accept the new setting by pressing the Select key.
To exit the service mode press the Mode key for 2 seconds.
Parameter adjustment up -key
21
0
Parameter adjustment keys
Select key

Table of Contents

Other manuals for Planmeca ProX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProX and is the answer not in the manual?

Planmeca ProX Specifications

General IconGeneral
TypeIntraoral X-ray
Tube current7 mA
Imaging TechnologyDigital
Focal spot0.4 mm
X-ray generatorHigh-frequency
Power Requirements100–240 V AC, 50/60 Hz
Tube voltage60-70 kV

Related product manuals