4 SERVICE MODE
26 Planmeca ProX Technical manual
After the exposure counter is shown on the display it can be set to zero by pressing and
holding down the ENDODONTIC / BW key for 5 seconds.
The counter is automatically set to zero after 50000 exposures.
Accept the display mode by pressing the Select key.
To exit the service mode press the Mode key for 2 seconds.
4.13 DAP display mode
Enter the service mode according to the instruction given in section 4.2 “How to enter/exit
service mode” on page 19.
Press the parameter adjustment up key until the parameter number 24 appears on the kV
display.
Press and hold down the Select key until the number starts to blink. The number can now
be changed with the parameter adjustment keys.
0: OFF
1: DAP display after exposure
2: DAP display using MODE key
3: 1+2
Accept the new DAP mode by pressing the Select key.
To exit the service mode press the Mode key for 2 seconds.
Parameter adjustment up -key
24
Parameter adjustment keys
Select key