4 SERVICE MODE
28 Planmeca ProX Technical manual
4.15 Maximum mA value
Enter the service mode according to the instruction given in section 4.2 “How to enter/exit
service mode” on page 19.
Press the parameter adjustment up key until the parameter number 26 appears on the kV
display.
Press and hold down the Select key until the number starts to blink. The number can now
be changed with the parameter adjustment keys.
0: maximum mA value 6 mA
1: maximum mA value 8 mA
Accept the new maximum mA value by pressing the Select key.
To exit the service mode press the Mode key for 2 seconds.
Parameter adjustment up -key
26
Parameter adjustment keys
Select key
0