5 RECALLING FACTORY PREPROGRAMMED EXPOSURE VALUES
Technical manual Planmeca ProX 29
5 RECALLING FACTORY PREPROGRAMMED
EXPOSURE VALUES
The factory preprogrammed exposure values can be found in the Planmeca ProX User’s
manual, section "EXPOSURE VALUES".
Press and hold down any of the preprogrammed setting keys when switching the unit on.
The error code E.29 will appear on the time display.
Do not clear the error code by pressing the select key, but press the Occlusal exposure
key for 6 seconds. The factory preprogrammed values and density value 0 will be stored
into memory.
E.29