7 TROUBLESHOOTING
Technical manual Planmeca ProX 35
7 TROUBLESHOOTING
The control panel displays do not come on
1. Problems in mains voltage, or a fuse is blown.
Check the mains voltage. Check the mains cable. Check and replace, if necessary, the
fuses located on the lower left side of the generator box. Open the generator box and
check and replace, if necessary, the generator fuse F2. Replace the generator. (Perform
the checks and parts replacements in this order).
2. Control panel power failure, or the control panel is defective. The control panel operates
with 12 V produced on the generator low voltage supply.
Check the control panel cable and the telephone cable. Replace the control panel.
(Perform the checks and parts replacements in this order).
Temperature of the tube head too high
If the temperature of the tube head exceeds 50 °C, the temperature will appear on the time
display. The control panel does not operate. Wait until the temperature drops.
50 C

Table of Contents

Other manuals for Planmeca ProX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProX and is the answer not in the manual?

Planmeca ProX Specifications

General IconGeneral
TypeIntraoral X-ray
Tube current7 mA
Imaging TechnologyDigital
Focal spot0.4 mm
X-ray generatorHigh-frequency
Power Requirements100–240 V AC, 50/60 Hz
Tube voltage60-70 kV

Related product manuals