4 SERVICE MODE
Technical manual Planmeca ProX 25
4.11 mA minimum value selection
Enter the service mode according to the instruction given in section 4.2 “How to enter/exit
service mode” on page 19.
Press the parameter adjustment up key until the parameter number 22 appears on the kV
display
The minimum mA value (2 - 6) is shown on the time display.
Press the Select key until the mA minimum value starts to blink. The minimum value can
now be changed with the parameter adjustment keys.
Accept the new mA minimum value by pressing the Select key.
To exit the service mode press the Mode key for 2 seconds.
4.12 Exposure count display mode
Enter the service mode according to the instruction given in section 4.2 “How to enter/exit
service mode” on page 19.
Press the parameter adjustment up key until the parameter number 23 appears on the kV
display.
1: Display is in Service mode.
2: Display is in both Normal operating mode / Service mode.
Press and hold down the Select key until the number starts to blink. The number can now
be changed with the parameter adjustment keys.
To show the exposure counter (EC 00000 - 50000)on the
mA
and
s
display press briefly
the MODE key.
Parameter adjustment up -key
22
2
Parameter adjustment keys
Select key
Parameter adjustment up -key
23
1
Parameter adjustment keys
Select key

Table of Contents

Other manuals for Planmeca ProX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProX and is the answer not in the manual?

Planmeca ProX Specifications

General IconGeneral
TypeIntraoral X-ray
Tube current7 mA
Imaging TechnologyDigital
Focal spot0.4 mm
X-ray generatorHigh-frequency
Power Requirements100–240 V AC, 50/60 Hz
Tube voltage60-70 kV

Related product manuals