3 EMC INFORMATION
18 Planmeca ProX Technical manual
Recommended separation distances between
portable and mobile RF communications equipment and Planmeca ProX X-ray unit
Planmeca ProX intraoral X-ray unit is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of Planmeca ProX intraoral X-ray unit
can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the Planmeca ProX intraoral X-ray unit as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800
MHz
800 MHz to 2.5
GHz
0.01 0.2 0.2 0.3
0.1 0.4 0.4 0.7
11.21.22.4
10 4.0 4.0 8.0
100 12.0 12.0 24.0