The System is grounded through the grounding conductor inside the power cable. This protective
grounding is essential for safe operation.
Verify fan operation when operating the device and stop device operation when fan is not operating.
Verify that once the Foot Switch is released, the treatment will stop and the counter on touch screen
will stop as well
2.6 Fire Hazards
Do not use the System in the presence of explosive or highly flammable materials like alcohol, methanol,
acetone, etc. (The suggested use of the Glycerin for TriPollar® treatments is in no way an explosive or
highly flammable material and is approved for usage).
Do not have flammable substances in the treatment room when preparing the skin for treatment.
If alcohol is used for cleaning, it must be allowed to dry thoroughly before the System is used. The
container of alcohol must be moved away from the System.
3. Safety Features:
The System is designed for safe and reliable treatment when used in accordance with proper
operation and maintenance procedures. The operator and all other personnel operating or
maintaining the System should be familiar with the safety information provided in this chapter.
• All Applicators must be connected to the system and locked prior to activation.
• To start the treatment, the Foot Switch must be pressed down by the operator. Output is emitted
only after pressing the Foot Switch.
• No RF signal is delivered to the VO Applicator unless a treatment Tip is connected to the
Applicator.
• The VO Tip electrode matrix in contact with tissue is pre-sterilized and designed for single use to
avoid cross contamination.
• Patient-Controlled Manual Switch of sensation automatically terminates application of RF energy
when pressed.
• The system automatically identifies the type of tip connected.
• An independent electronic circuit stops the operation of the system in case of a software error.
• After each press and release of the Foot Switch, the LED indicator guide ring turns Green. LED
Indicator guide ring will turn Orange when the Foot Switch is pressed and treatment pulse power
is transmitted to the electrodes. After treatment termination, the LED indicator guide ring will
turn green and ONLY then, the operator can release the foot switch.