Proprietary Information of Pollogen a company of Lumenis 32
10.3.5 Attach VO disposable tips
To connect the VO disposable tips, gen 36, gen36L,
gen100, gen 12 or H7X7: do the following:
Select a blister pack containing from one of
the above mentioned tip boxes.
Verify valid expiration date printed on the
tip’s blister pack. (gen 36 is provided as an
example)
Remove cover from the blister pack.
Place on a flat surface
Apply sterile gloves
Connect the tip to the VO Applicator. See
detailed instructions below
To connect the tip, align the raised strip on
the tip with the groove on the applicator
and push them together
Push in and turn the tip clockwise to lock it
in place Verify that the tip is locked onto
the applicator’s head
Remove tip cap
To disconnect the tip, turn the tip
counterclockwise
After the treatment, discard the used tip
with biological hazard materials
10.3.6 Connect and lock Applicator
No. 1
Make sure it is properly attached and
fastened securely.