Legend Pro System User Manual Version 7
Proprietary Information of Pollogen a company of Lumenis 51
the treated area
Suggested number of treatments may vary between individuals and between treatment areas
Typical course of VoluDerm ™ treatment involves 4-5 treatments: 3 treatments spaced 1-2 weeks
apart (depends on skin condition and healing rate) followed by additional 1-2 treatments spaced 2-3
weeks apart
Before performing the next treatment, it is important to always verify that the treatment area has
completely healed.
Patient medical history should be updated before each session including sun exposure, new
medications etc.
A touch up treatment may be needed approximately every six months, according to individual need
and response
12.2 TriPollar® Energy Treatments using Applicators 1, 2, 3
12.2.1 Pretreatment guidelines:
Patient safety can be ensured only when well trained personnel operate the system
A patient’s medical history should be completed prior to scheduling a treatment
Never treat a patient with any of the contraindications criteria listed (see section 4)
Patients should be fully informed of the treatment protocols including inclusion and exclusion criteria,
clear expectations/realistic results and if there are any risks associated with the treatment
12.2.2 During the treatment
When placing the applicator on the skin, make sure it is not transmitting RF power (Foot switch is not
pressed and Indication ring light on the applicator is green)
Make sure to move the applicator at all times on skin surface, never stop moving the applicator on
the skin (even when it is not transmitting RF power and the indicator light is green)
Using power that is higher than needed or longer treatment sessions can cause over-treatment and
lead to undesired over-heating of the skin
The desired end point surface temperature for facial treatments should range between 40-42
o
C (104-
107.6
o
F). Never exceed 43
o
C (109.4
o
F) when treating the face
When treating a patient for the first time, always begin treatment with COURSE A default settings for
RF power
Do not increase power level at once by more than 2 increments
Exact treatment parameters are individual and should be adjusted (increased or decreased) according
to patient feedback and skin tolerance
Upon reaching the endpoint skin surface temperature, treatment should be continued for at least the
recommended exposure time to maintain the generated heat
In the case of patients with sensitive skin, it is recommended to gradually reduce power level.
Maximum power level for Applicator 3 is restricted to 30
During the treatment, make sure to monitor the surface temperature to ensure power settings are
suitable
Never increase surface skin temperature over 43
o
C (109.4
o
F)
NOTE: When treating facial area, never treat the area below the eyebrow & never treat over
the eye lids & eyeballs.