Legend Pro™ System User Manual Version 7
Proprietary Information of Pollogen a company of Lumenis 40
11.4.3 VO Applicator tips
If a treatment tip is not connected, a warning message is
displayed.
The system does not enable starting a VoluDerm ™
treatment until a tip is connected. VoluDerm ™ power will
not be emitted unless a tip is connected (For treatment tip
connection instructions, please refer to “To connect the
tip”).
During the treatment, all buttons will be dimmed/not active.
During the treatment, if the tip is removed, a warning
message is displayed.
The system does not enable to proceed with VoluDerm ™
treatment until a tip is re-connected.
Once the tip is connected to the VO Applicator, verify
the type of tip indicated on the screen before starting
the treatment.
The system shall automatically identify the type of tip
attached to the VO applicator and select treatment
parameters according to tip selected.
During the treatment, if the tip is removed, and another
type of tip is connected, a warning message is displayed.
The system does not enable to proceed with VoluDerm ™
treatment until the same type of tip is connected.
This message will also be applicable for gen100, gen 12 and
H7X7 tips and gen36L.
VO tip pulse limit is 800 pulses. When starting a new
treatment using VO tip with more than 800 pulses, the
system will show 'tip is used' screen.
In order to continue, please replace the tip with a new one.