EasyManua.ls Logo

POLLOGEN Legend Pro - General overview and Installation Requirements; Preparation for Operation

POLLOGEN Legend Pro
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Legend Pro System User Manual Version 7
Proprietary Information of Pollogen a company of Lumenis 29
10.2 General overview and Installation Requirements
Electrical Requirements
The System has a universal power inlet Single phase 100 - 240 VAC; 5A; 50-60Hz.
The system must be connected to a grounded power outlet (3-pin). Only supplied cable is to be used.
(Cable 10A, 250V, 1.8m length)
Environmental Requirements
Corrosive materials can damage electronic parts. Therefore, the System should only be operated in a
non-corrosive atmosphere. For optimal operation of the System, maintain room temperature between
15º-30ºC and keep relative humidity at less than 80% and maintain Atmospheric pressure of 70kPa to
106kPa.
Moving the System
1. Place the Applicators into their cradles.
2. Turn the System off.
3. Disconnect the power cord, Foot Switch and Patient-Controlled Manual Switch.
4. Release the wheel brakes.
5. With one hand securing the Applicator, slowly pull the System using its handle.
C A U T I O N
ONLY MOVE THE SYSTEM BY USING ITS HANDLE. PUSHING OR PULLING THE SYSTEM
WITHOUT USING ITS HANDLE MAY TURN THE SYSTEM OVER AND RESULT IN
DAMAGE TO THE SYSTEM.
10.3 Preparation for Operation
10.3.1 Applicator and cable holder’s assembly procedure:
Take both applicators holders and screws
out of the box

Table of Contents