Legend Pro™ System User Manual Version 7
Proprietary Information of Pollogen a company of Lumenis 48
12.1.3.1 VO : Default Treatment parameters:
VoluDerm ™: Default Treatment parameters:
* Values in parenthesis represent the range within each level
VoluDerm ™ notes:
You may adjust VoluDerm ™ Power & Exposure parameters within the indicated range
To adjust the VoluDerm ™ output, start by adjusting the % Power. Change of % Exposure is only for
advanced usage
Mild vibration is felt during the initiation of VO pulse
Always wait for the BEEP sound before moving to the adjacent pulse. Do not move the Applicator until
you hear a BEEP sound.
TF H7X7 Notes:
You may adjust RF Exposure parameters within the indicated range
Make sure the skin is completely dry and that there is full contact of all pins with the skin, ensure it
by stretching the skin and applying mild pressure.
12.1.4 Treatment Procedure
Selecting the appropriate program and tip (gen 12, gen 36, gen36L , gen100 or H7X7) is essential for a
successful treatment
The choice of treatment parameters depends on the treatment area, skin condition, patient skin type,
and tolerance
For each tip there are three levels: Low, Medium and High. It is recommended to always start with the