12.2.3 Treatment Parameters
The following table shows the default treatment parameters based on physician’s experience with the
TriPollar® Applicators.
Table 5: Treatment Parameters for TriPollar®
Treatment parameters vary between individuals, affected by speed of applicator movement and movement
styles. It is important to begin systematically with default parameters when treating a patient for the first
time.
Always start a new patient with Course A default parameters. You may use course B in further treatments.
The below table represents treatment values that are based on the range of RF power used for face
treatment in clinical trials.
Table 6: Typical treatment parameters for face treatment
NOTE:
Always use default RF power settings when treating a patient for the first time
It is important to define your treatment settings by following the treatment instructions in Section
9.2.3
12.2.4 TriPollar® treatment procedure
Pre-Treatment Preparation
During the first visit, the operator or an authorized staff member should assess the areas intended for
treatment and take the following actions:
Take a detailed patient medical and physical history including previous treatment methods
Determine if the patient is suitable for treatment - do not treat patients with any of the
contraindications outlined in this User Manual
Determine why the patient is seeking treatment and what his/her expectations are
Inform the patient as to the course of treatment and possible side effects
Instruct the patient as to the safety warnings
Instruct the patient to remove all jewelry prior to treatment