Legend Pro System User Manual Version 7
Proprietary Information of Pollogen a company of Lumenis 19
7.1.11 Non-Contact Infrared
Thermometer:
Figure 15: Non-Contact Infrared Thermometer
The Infrared thermometer (Figure 15: Non-Contact Infrared
Thermometer), measures the surface temperature of the skin.
An optical component of the unit collects energy and focuses
it onto a detector. Then an electronic component translates
the information into a temperature reading which is displayed
on the thermometer.
The laser is used for aiming purpose only.
Please note that calibration is required yearly and is within your
responsibilities.
Lumenis does not calibrate or follow up thermometer
calibration.
8. Technical Specifications
Parameter
Value/Data
Input voltage
100-240 Volt, 50-60Hz, max 2.2A
Mode of operation
VoluDerm, TriFractional, TriPollar
Maximum output power RF
50 Watts @ 200 Ohm
Maximum output voltage
100±10 Vrms
Output frequency RF
1MHz
Output power control RF
Pulse width modulation (PWM)
Weight
~66 Kgs
Dimensions
L1.5 xW1.5 x H3.6 Inch
Environmental Conditions for Operation
Temperature
15°C to +30°C (59°F to 86°F )
Relative humidity
Up to 80%; non-condensing
Atmospheric pressure
70-106 kPa
Maximum Altitude
2000 meters
Environmental Conditions for Transportation & Storage
Temperature
-20°C to +55°C (-4°F to 131°F)
Relative humidity
80%; non-condensing
Atmospheric pressure
50-106 kPa
Maximum Altitude
2000 meters
Table 2: Technical Specifications
IEC 60601-1 Classification Class I